The DLM-080 FUTURE Digital Recorder – User’s Manual
26.11.2004
32
time-meter (F44)
F69
0 0..2
Procedure of operation for CHANNEL 8 defined as
time-meter (F45)
as before
F70
0 0..255
Allowable difference between set value and read out
value for channel 1. Alarm will indicate cases,
when this difference is exceeded.
F71
0 0..255
Allowable difference between set value and read out
value for channel 2. Alarm will indicate cases,
when this difference is exceeded.
F72
0 0..255
Allowable difference between set value and read out
value for channel 3. Alarm will indicate cases,
when this difference is exceeded.
F73
0 0..255
Allowable difference between set value and read out
value for channel 4. Alarm will indicate cases,
when this difference is exceeded.
F74
0 0..255
Allowable difference between set value and read out
value for channel 5. Alarm will indicate cases,
when this difference is exceeded.
F75
0 0..255
Allowable difference between set value and read out
value for channel 6. Alarm will indicate cases,
when this difference is exceeded.
F76
0 0..255
Allowable difference between set value and read out
value for channel 7. Alarm will indicate cases,
when this difference is exceeded.
F77
0 0..255
Allowable difference between set value and read out
value for channel 8. Alarm will indicate cases,
when this difference is exceeded.
F78
Free
F79
Free
F80
0 0..1
PT-100 measurement range
0 – range up to 400°C
1 – range up to 600°C
F81
0 0..1
Recording base 0 – seconds
1 – minutes
F82
0 0..1
Digital filter for the AC converter on / off 0 – filter off
1 – filter on
F83
1 0..1
Transmission speed 0 – 9600 bps
1 – 19200 bps
(THE ONLY TRANSMISSION
SPEED AVAILABLE FOR
MODBUS RTU PROTOCOL IS
9600; SET PROPER VALUE
)
F84
0 0/1
Printing base 0 – seconds
1 – minutes
F85
0 0..1
Clock / time-meter
“0”- displays real-time clock
“1” – displays impulse counter
on the 0/1 24V DC input
F86
0 0..1
Relay status – REL 9 “0” – relay as an alarm
signalling device